TWP Weighing Indicator Service Manual
- 24 -
OL or( -------) appear the display:
Maximum capacity exceeded.
Power supply faulty. Check all power cables and connecters.
Calibrate again with correct calibration weights.
Check load cell connections.
Load cell damaged.
(------) or Lo:
Weight is below permissible limit.
Check the pan installed correctly.
Calibrate again with correct calibration weights.
Power Supply faulty. Check all power cables and connecters.
Load cell damaged. Check load cell connections and connecters.
Try to turn on again.
Unstable display:
Check the pan is seated proper and touching some where
Check any vibrations, noises, sudden temperature changes
Check power supply.
Check battery and adaptor, connect to plug for charging.
Check the load weight is seated properly.
Check the load cell connections and connecters.
Incorrect value:
Calibration error.
Calibrate again with exact calibration mass weights.
Check item if it is on the platter properly and avoid touching the surface.
Check power supply and battery.
Check load cell connections and connectors