TWP Weighing Indicator Service Manual
- 23 -
8. MAINTENANCE
CLEANING
Disconnect the power before cleaning.
Use a cloth with mild suds and light cleaning agents. Make sure that fluid not
able to get into the device. Use a clean and soft cloth for rub off.
ERROR CODES
Error Message
Description
Solution
-----
Maximum load exceeded
Unload or reduce weight
Err 4
Zero setting error
Zero setting range exceeded
due to switching on.(4%max)
Make sure platform empty.
Err 6
A/D value out of the
range
Make sure platform empty and
check the pan is installed
proper. Check the load cell
connectors.
TROUBLE SHOOTING
No Display:
Mains power is turned off or power supply not plugged proper.
Power supply faulty.
Internal Battery is not charged.
Check On/OFF switch is ok or not.
Check the PCB power connections and cable.
Display is Blank after the self test / Err stuck:
Unstable weight.
Check the platform is installed correctly.
Try again to turning on.
Check the load cell is not touching any where.
Load cell is damaged. Check the load cell connections and all.