Appendix C: Field Sensor Replacement
The DC710 uses field replaceable smart sensors. When you receive a sensor
it will come pre-calibrated. Factory calibration once per year is still impor-
tant to ensure your analyzer is functioning properly on all functions and
ranges.
Contact your TPI distributor to purchase replacement sensors.
Note: Y
ou must wait 24 hours after sensor replacement before using your
analzyer.
Follow these steps to replace the sensors:
1. Remove the protective rubber end caps using your hands to pry them from
both ends of the analyzer. .
2. Locate the four screws holding the housing together. (see picture below)
3. Loosen the screw and remove the back housing by separating it from the
front housing.
25