Introduction
Thank you for purchasing TPI brand products. The TPI DC710 com-
bustion analyzer is state of the art and easy to use. Connect it to your
smart device and you are ready to perform combustion analysis. The
instrument is ruggedly constructed and comes with a 3 Year unit and 2
Year sensor Guarantee.
Download the TPI View app from the Apple
®
app store or the Google
®
play store to begin using your DC710.
General Overview
The TPI DC710 combustion analyzer uses field replaceable state of the
art electrochemical sensors. This sensor technology provides the
longest lasting most accurate and reliable means for performing com-
bustion tests. The sensors in your analyzer will need to be replaced
periodically and calibration is recommended once every year.
Electrochemical sensors by nature are always active. Therefore the
time the analyzer is off and not being used must be taken into account
when determining sensor life. The sensors in your analyzer are war-
ranted for two years. This warranty does not cover sensors damaged
through misuse of the analyzer.
You should keep the battery of your DC710 charged so power is con-
stantly being supplied to your sensors.
The following guidelines will help prevent damage to your sensors:
Always use the mini pump filter when testing flue gases.
Always periodically check and replace the mini pump filter as needed.
Always make sure the in-line filter / water trap is installed properly.
Always periodically check and replace the in-line filter as needed.
Always remove water or condensation from the inside of the in-line fil-
ter / water trap assembly prior to performing tests.
Always use the optional oil filter (p/n A773) when performing tests on
oil burning equipment.
1