EasyManua.ls Logo

TPS WP-80 - Determining if an instrument or electrode is faulty

TPS WP-80
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18.1.3 Temperature Compensation
The slope of a pH electrode (section 18.1.2) is affected by temperature.
This effect is compensated for either by using an Automatic Temperature
Compensation (ATC) probe or by entering the sample temperature
manually. Figure 18-3 shows the slope of a pH electrode at various
temperatures.
-600
-400
-200
0
200
400
600
0 7 14
mV
Electrode
Potential (mV) at
0 oC (54mV/pH)
Electrode
Potential (mV) at
50 oC (64mV/pH)
Electrode
Potential (mV) at
100 oC
(74mV/pH)
pH Electrode Response, as a Function of Temperature
pH
Figure 18-3
18.2 Checking the reference junction of a pH electrode.
If pH readings are inaccurate or unstable, the reference junction of the
electrode may be blocked. The following test can be performed to
determine if the reference junction of a pH electrode is making adequate
contact with the sample solution.
1. Calibrate the WP-80, as per section 4.1.
2. Dilute 1 part of pH6.88 buffer with 9 parts of distilled water.
3. Measure the pH of the diluted buffer. The result should be 7.06 +/-0.02
pH.
4. If the value obtained is outside of these limits, then clean the reference
junction, as per the instructions supplied with the pH electrode.
5. Re-calibrate the WP-80 and repeat the test.
6. If the value obtained is still outside 7.06 +/-0.02 pH, then the electrode
should be replaced.

Related product manuals