EasyManuals Logo

Transcend Micro User Manual

Transcend Micro
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
Page 16
1. Remove the HME assembly from your mask
2. Remove the HME assembly from your hose
3. Unscrew the HME adapter bottom from the HME adapter top.
4. Discard the HME cartridge. The HME cartridge plastic and foam are not to be
cleaned and should only be replaced.
5. After following cleaning process, refer to steps in the Connecting Comfort Feature
section to reassemble the HME adapter.
Cleaning the Exterior
Follow these instructions to clean the exterior of the Transcend Micro device.
1. Mix a solution of 5% mild liquid detergent in distilled water (1.6 fl. Oz. liquid
detergent per quart of distilled water). Mild detergent should contain biodegradable
anionic surfactants and no phosphate.
2. Submerge a lint-free cotton cloth into the detergent solution.
3. Wring excess water from the cloth then wipe the exterior surface of the Transcend
Micro device for approximately 20 seconds using a gentle back and forth wiping
motion. Ensure contact with all accessible surfaces to adequately remove any soil
buildup.
4. Rinse the cloth in clear water to remove residual cleaning solution.
5. Wring excess water from the cloth and wipe the Transcend Micro using a gentle
front to back wiping motion to remove any detergent solution remaining on the
surface.
6. Wipe the device with a dry, lint-free cotton cloth until the device is fully dry.
Cleaning of Accessories
The following accessories should be cleaned with a 5% solution of mild dishwashing liquid
detergent in distilled water (1.6 fl. oz. liquid cleaning detergent per quart of distilled
water). Mild detergent should contain biodegradable anionic surfactants and no
phosphate.
Clean the device and its components as shown in the schedules below in order to maintain
the quality of your device and to help prevent the growth of germs that can adversely
impact your health.
Accessory
Periodic Cleaning Cycle
Product Service Life
AirFlex Hose
Weekly
3-Month
WhisperSoft Muffler
Weekly
3-Month
AirMist HME
Weekly
3-Month

Table of Contents

Other manuals for Transcend Micro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Transcend Micro and is the answer not in the manual?

Transcend Micro Specifications

General IconGeneral
BrandTranscend
ProductMicro
Pressure Range4-20 cmH2O
Altitude AdjustmentAutomatic
Operating ModesCPAP

Related product manuals