Electromagnetic Compatibility
The Transcend Micro complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2:2014 for residential, commercial and light
industry environments. Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables, than the recommended distance
of separation (30CM). The Transcend Micro has been designed to meet EMC standards,
however should you suspect the device performance is being affected by other equipment,
move the device away from the potential cause of the interference. The Transcend Micro
complies with Part 15 of the FCC rules and Industry Canada license-exempt RSS
standards. Operation is subject to the following two conditions 1) this device may not
cause harmful interference, and 2) this device must accept any interference received,
including interference that may cause undesired operation.
Electromagnetic Emissions
The Transcend Micro is intended for use in the electromagnetic environment specified
below. The customer or the user of the system should ensure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environmentguidance
RF radiated
emissions
CISPR 11
Group 1
The Transcend Micro uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF conducted
emissions
CISPR 11
Class B
The Transcend Micro is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic
distortion
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Complies

Table of Contents

Other manuals for Transcend Micro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Transcend Micro and is the answer not in the manual?

Transcend Micro Specifications

General IconGeneral
BrandTranscend
ProductMicro
Pressure Range4-20 cmH2O
Altitude AdjustmentAutomatic
Operating ModesCPAP

Related product manuals