Technical Specifications
Transcend Micro weight:
0.48 lbs (219.5g)
Transcend Micro dimensions:
3.6 in x 3.6“ x 2.4“ (9.1 cm x 9.1 cm x 6.1 cm)
Air outlet connector port
dimensions:
Custom interface to device, standard 22-mm
diameter connection to user interface
Design life
5 years
AC Power Supply PSA 5
AC supply input:
100-240 VAC, 50-60Hz, 1.0A Max
AC supply output:
19 VDC 2.1 Amp
Battery (optional)
Transcend P8 Battery
14.4 VDC, 5,200 mAH
Transcend Micro Performance
Working pressure range:
4 to 20 cm H
2
O
Accuracy of pressure setting:
±1 cm H
2
O or ±10%, whichever is greater
Maximum system shutdown
pressure:
30 cm H
2
O
Ramp time duration:
0-45 min + 25% time variance
Operating temperature range:
41 to 95°F (5 to 35°C)
Storage/transport temperature
range:
-4 to 140ºF (-20 to 60ºC)
Operating humidity range:
10% to 80% relative humidity, non-condensing
Storage/transport humidity
range:
10% to 90% relative humidity, non-condensing
Altitude range:
0-8000 feet (Automatically adjusted)
NOTE: Allow one hour for the device to either warm or cool when coming from
storage temperature to operating temperature prior to use