Transcend Micro - Airflow and LED Indicator Problems

Transcend Micro
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 20
Dry nose or throat
irritation.
Dry air
Add humidity to the room.
Replace HME if being utilized.
Secure Transcend AirMist HME.
Contact your homecare provider.
Dirty air path accessory
Follow cleaning instructions for
accessories.
Dirty air filter
Change the air filter.
Device control panel
LEDs don’t flash or
illuminate when the
power supply is
connected to the
device.
Power source is not
properly connected.
Check all power connections and
verify if the LED light on the
power supply is illuminated.
AC power may not be
active.
Use another power outlet.
Confirm outlet is not controlled
by a wall switch.
No airflow from the
device.
Device motor failure; or,
electronics failure
Contact the homecare provider’s
technical service department.
All LEDs flash two
times then pause,
then flash again
Device detects an
operating error
Hold down the power button until
all LEDs stops flashing.
Release the power button, the
device will reset and power up in
Standby Mode.
Check hose, mask and any
accessories in the breathing
circuit for air leaks and proper fit.
If problems persist, contact your
homecare provider
Device shuts down
during therapy
Improper seal of external
hardware (mask, tubing);
or use of external
hardware past
recommended service
life.
Verify all external equipment in
breathing circuit is seated
correctly to ensure a proper seal.
Check hose for leaks.
Follow mask manufacturer
recommendations for fit and seal.