Dry nose or throat
irritation.
Add humidity to the room.
Replace HME if being utilized.
Secure Transcend AirMist HME.
Contact your homecare provider.
Follow cleaning instructions for
accessories.
Device control panel
LEDs don’t flash or
illuminate when the
power supply is
connected to the
device.
Power source is not
properly connected.
Check all power connections and
verify if the LED light on the
power supply is illuminated.
AC power may not be
active.
Use another power outlet.
Confirm outlet is not controlled
by a wall switch.
No airflow from the
device.
Device motor failure; or,
electronics failure
Contact the homecare provider’s
technical service department.
All LEDs flash two
times then pause,
then flash again
Device detects an
operating error
Hold down the power button until
all LEDs stops flashing.
Release the power button, the
device will reset and power up in
Standby Mode.
Check hose, mask and any
accessories in the breathing
circuit for air leaks and proper fit.
If problems persist, contact your
homecare provider
Device shuts down
during therapy
•
Improper seal of external
hardware (mask, tubing);
or use of external
hardware past
recommended service
life.
Verify all external equipment in
breathing circuit is seated
correctly to ensure a proper seal.
Check hose for leaks.
Follow mask manufacturer
recommendations for fit and seal.