- 16 -
Option board dependent menu items
SEr 1. SEr 2. SEr 3. SEr 4. _Addr
Menu items related to serial communications. These will only appear if an RS232,
RS485, or USB I/O board is detected. If so, see Section 16.
The Cal Instrument indicator starts flashing 30 days before instrument
calibration is due. To acknowledge and turn off flashing, change the first
digit of this menu item from 1 to 0. The Cal Sensor indicator starts
flashing 30 days before sensor calibration is due. To acknowledge and turn
off flashing, change the second digit of this menu item from 1 to 0.
Use keypad to enter the current date in MM.DD.YY format. This menu item
can also appear as an error message when the battery has just been
replaced, or in other instances where the instrument date is not correct.
The meter remains non-functional until the current date has been entered.
00.00.00 Meter calibration date in MM.DD.YY format. Read-only.
00.00.00 Sensor calibration date in MM.DD.YY format. Read-only.
SEr no
Transducer
Serial No.
If the serial number is 6 digits or less, press the Peak key once to display
all digits 123456. If the serial number is greater than 6 digits, press Peak
once to display the first digits 12–___, and press the Peak key again to
display the remaining digits 345678. Read only.
UnitS_
Units of mea-
sure. Read
only.
_A_ b_C Initials of the person who performed the transducer calibration.
Read only. Characters used are:
___365 Required transducer calibration interval in days. Read only.
brdg-A or brdg-b or brdg-C Identifies whether meter is monitoring
bridge A, bridge B, or bridge C.