4–12
Enter any desired information (up to eight characters) and
press
ENTER/START. The gauge stores the last displayed
results and notes under the active project and returns to the
Ready screen. For more information on storing measured
data, see the Storing Data section on page 5-10.
NOTE
The gauge stores only the data shown in the last display. It
does not store the results of each individual reading.
Checking the Gauge during Use
To verify that the gauge is reading properly during use,
perform the following check:
1. Ensure that the gauge is turned on and in the correct
measurement mode.
2. Turn off (disable) all operator-installed density offsets and
mix calibrations. See Chapter 3 for more information.
3. Place the gauge on the reference standard and verify that
it reads within ± 0.5 pcf (±8 kg/m
3
) of the actual density
value printed on the gray label located inside the case lid.
If the reading is outside these limits, take a reference
reading as described above to adjust the gauge.
Note that each reference standard is unique to that individual
gauge. Use care to not intermix different gauges and cases.
Recall
To view the results of the most recent measurement, press the
RECALL key from the Ready screen. The gauge displays the
results of the most recent measurement. To return to the
Ready screen, press the
ENTER/START key.