4-1
CHAPTER 4
Operating SMPS
Hardware
After you have set up the SMPS spectrometer plumbing in
Chapter 3
, use the information in this chapter to do the following:
Select flow rates
Set flow rates
Check the SMPS
spectrometer for laminar airflow
Before You Begin
Power should be applied to the Condensation Particle Counter
(CPC) and the Electrostatic Classifier. The analog output must be
set to HOST in the Model 3010, 3022A, or 3025A CPC. See
Applying Power Sequence
” in Chapter 3 for more information. On
the Classifier, the voltage control should be set to “Analog Ctrl”
(lower left pane on displaysee Figure 4-1) to allow the DMA
voltage to be controlled by the CPC.
Figure 4-1
Classifier LCD Display