4-2 Model 3936 Scanning Mobility Particle Sizer
TM
(SMPS
TM
) Spectrometer
Selecting Flow Rates
To run the Aerosol Instrument Manager
®
for SMPS™ software, the
ratio of the Sheath Air to Polydisperse Aerosol flow rates on the
Electrostatic Classifier is normally set to 10:1. To achieve the best
resolution, the Sheath flow rate and the Excess flow rate must
equal each other, and also the Polydisperse Aerosol flow rate and
the Monodisperse Aerosol flow rate must also equal each other.
Figure 4-2
Schematic Flow Diagram of the Classifier with Long DMA
With the Model 3080 Classifier, flow control is very easy (see
Figure 4-2 and Figure 4-3). The Sheath and Excess flows are
running in a continuous loop so they have to be equal. The
polydisperse flow is controlled by the CPC drawing aerosol through
the Classifier, so the monodisperse and polydisperse flows have to
be equal. The Sheath flow is controlled by setting the appropriate
value on the front panel of the Classifier. The CPC flow is fixed at
1 L/min, in the case of a 3772, 3010 CPC or 3785 WCPC. For a
3782 WCPC or 3786 UWCPC, the flow is fixed at 0.6 L/min. And
the flow can be toggled between 0.3 L/min or 1.5 L/min in the case
of the 3775, 3776, 3022A or 3025A CPC. Other flows require the