Operation
29
Note
Greater accuracy will be obtained with longer samples. The time
you permit for sampling often depends on the reference
instrument and characteristics of the measured aerosol. It may
take some time to collect sufficient aerosol onto a filter cassette
for accurate gravimetric analysis. Refer to instructions of your
reference instrument for sampling times.
6. Stop sampling with both instruments at the same time.
7. Record the DustTrak monitor average concentration by viewing the
sample average in the Data screen. (Data Screen is reviewed later in this
chapter.)
8. Determine the mass concentration in mg/m
3
from your reference
instrument. For gravimetric sampling this means weighing the gravimetric
sample.
Note
If you used the internal gravimetric filter in the DustTrak Model
8530, the flow rate used to compute the concentration should be
2 L/min, not 3 L/min since only 2 L/min of aerosol flow reaches
the filter.
9. Compute the new calibration constant, NewCal, using the following
formula:
CurrentCal
ionConcentrat DustTrak
ionConcentrat Reference
NewCal
10. Select Photometric from the User Cal drop down selection and enter the
NewCal factor using the onscreen controls.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments dusttrak II 8530 and is the answer not in the manual?

TSI Instruments dusttrak II 8530 Specifications

General IconGeneral
BrandTSI Instruments
Modeldusttrak II 8530
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals