EasyManua.ls Logo

TSI Instruments SidePak AM520 - Page 48

TSI Instruments SidePak AM520
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3
32
Effect of Calibration Factors on Full Scale and Alarms
Note that the instrument maximum full scale mass
concentration reading is not fixed at 100 mg/m³; it changes
when the Photometric Calibration Factor is changed. The
same is true with the Response Factor and response
concentration reading.
For instance, if a Photometric Calibration Factor of 2.00 is
chosen, all readings are multiplied by 2.00, and the effective
maximum mass concentration reading of the instrument is then
200 mg/m³. Similarly, if the Response Factor is set at 25%, all
mass concentration readings are multiplied by .25, and the
effective maximum response concentration reading of the
instrument is 50 mg/m
3
.
Since the range of selectable Photometric Calibration Factors is
0.10 to 10.00, the effective maximum full scale mass
concentration reading can be anywhere between 10.0 and
1000 mg/m³.
The selectable range of 0% to 100% for the Response
Factor limits the maximum response concentration to
between 0.00 mg/m
3
and the current full scale mass
concentration reading.
Since all alarms must be allowed to approach full scale, the
maximum alarm value is 999 mg/m³. This creates a situation
where the alarm values can be set higher than the maximum
reading. In this case, the instrument will never be able to read
high enough to trigger the alarm.
To prevent this situation, the instrument displays the message
Warning: Alarm value is out of range” if you try to enter a
value that is out of range.

Table of Contents

Related product manuals