9.7 Replacing the Fuse
Caution
Make sure that the electrical power cord is disconnected!
Use a screwdriver to unlock the fuse holder cover by turning it counter
clockwise ¼ turn, and pull it out.
Insert a new fuse into the holder and turn the cover clockwise until
locked.
Make sure that the correct fuse is installed
1. Water Pump Fuse: 1.25 amps
2. Air Pump Fuse: 2.0 amps for 120V
1.25 amps for 230V
For autoclaves
Models E, EK
(without an air pump)
For autoclaves
Models EA
(with an air pump)