49
11. Fill the reservoir with distilled water or mineral free water only.
12. Press the manual water fill button and allow a small amount of water
(2-4 ounces) to fill chamber and flush out the fill tube. Remove water
from chamber.
13. The instrument is ready to use.
IMPORTANT:
DO NOT sterilize instruments during the cleaning process!!!
CAUTION:
Keep out of reach of children. Contains mildly acidic ingredients.
Avoid contact with the skin, eyes or clothing. Wash hands well
after touching the powder, in the case of eye contact flush with
continuous running water for at least 15 minutes. If irritation
persists get medical attention. If accidentally swallowed, do not
induce vomiting, drink large amounts of water and obtain
medical attention. MSDS available upon request.
Formodels 1730, 2340, 2540 use onepacket of
CHAMBER BRITE ™.
Formodels 3140, 3850, 3870 use twopackets of CHAMBER BRITE ™.
Clean every 20 cycles or as needed.