24
— Press the PROGRAM key to advance to the next
value.
— The cursor on the bottom row will move one
position to “R”; now enter the pressure value
recorded from the digital display to which you
have already added 14.5.
— Press the PROGRAM key one final time
If PROGRAM WAIT has not been displayed then
continue to press the PROGRAM key until it appears.
After a brief period the unit will return to a Ready
state
This completes the calibration procedure for
temperature e and pressure. A sterilization cycle
should be performed to verify that the system is
calibrated and running properly.