5.4.4 Accessing the calibration mode
With the unit turned on press the PROGRAM key.
— Pressing the PROGRAM key will bring you to its
first function of setting the system’s date the time.
Setting the date and time can be bypassed by pressing the
PROGRAM key six times.
On the seventh press of the PROGRAM key CODE:
∅∅∅ will be displayed.
Advance the code to 011 by pressing the UP arrow key
(this is a technician level code) then press the
PROGRAM key.
Continue pressing the program key until CALIB CODE:
1∅∅ is displayed.
— Using the UP arrow key select CALIB CODE: 107.
— Press the PROGRAM key again until Program Wait is
displayed, this is an intermediate calibration phase. Once
the program wait is completed TEMP CAL: ∅∅∅ will
be displayed.
— Using the UP arrow key advance to TEMP CAL: ∅∅1
— Press the PROGRAM key again. This is the Temperature
Calibration mode and the data is ready to be entered
— Press the PROGRAM key one additional time and the
display will show PRESS CAL: ∅∅∅.
— Using the UP arrow key advance to PRESS CAL: ∅∅1
— Press the PROGRAM key again. This is the Pressure
Calibration mode and the data is ready to be entered