8.2.19 Heat Step – sterile temperature overshoot control
This parameter is used to reduce overshoot of the sterilization
temperature by controlling the sterilization pressure. The
sterilization pressure is calculated by the unit based upon the
SteTemp parameter. When the cycle starts the heating rate in
the chamber is constant until the pressure in the chamber
reaches the sterilization pressure minus the Heat Step. From
this point on heating is performed in pulses. Each pulse lasts
15 seconds and increases the pressure in the chamber by a
small amount until the proper sterilization pressure is reached.
Access Code – 13
Resolution – 0.1 psi
Minimum Value – 0.1 psi
Maximum Value – 2 psi
Cycle
1 2 3 4 5 6
Default
Values
Value
0.7 0.7 0.7 0.7 0.7 0.4
8.2.20 PresInPsig – selects the pressure display units.
This parameter sets the units of pressure that are printed by the
printer and used on the digital display during St.By (Stand By)
and while running a cycle. The two choices are psia (absolute
pressure) or psig / InHg (gauge pressure and inches of
mercury) Changing this parameter does not effect the units of
calibration used in any other parameter.
Access Code – 13
Resolution – 1
Pressure is displayed in
Psig and vacuum in InHg – 1
Pressure is displayed in psia – 0
Cycle
1 2 3 4 5 6
Default
Values
Value
1 1 1 1 1 1
8.2.21 TempInC – selects the temperature units.
This parameter sets the units of temperature that are used by
the autoclave.
Access Code – 13
Resolution – 1
Temperature is displayed in °C – 1
Temperature is displayed in °F – 0
Cycle
1 2 3 4 5 6
Default
Values
Value
0 0 0 0 0 0