EasyManua.ls Logo

Tuttnauer 3870 EHS - Page 47

Tuttnauer 3870 EHS
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46
8.2.22 Flash – Heating over-pressure control
This parameter is designed to regulate over pressure during the
heat up stage. The system constantly monitors the temperature
and pressure and compares both values to the Saturated Steam
Curve. If the pressure is found to exceed the corresponding
temperature by more than 0.7 psi then the fast exhaust valve is
opened for 1 second in an attempt to realign the temperature
and pressure. The system will wait 15 seconds and check
temperature and pressure again. If needed the fast exhaust will
be opened again for 1 second. This process will continue until
the temperature and pressure are correctly aligned.
Access Code – 13
Resolution 1
Value 0 or 1
If Flash = 0 This feature is not activated.
If Flash = 1 This feature is activated.
Cycle
1 2 3 4 5 6
Default
Values
Value
1 1 1 1 0 0
8.2.23 Auto. Num – Autoclave Identification number
This parameter is used to set the identification number relative
to other autoclaves in the facility. This number only appears on
the printout of a unit with a printer installed. It enables the
operator to link the printout with a specific autoclave. This
parameter need only be set once, while in any of the 6
programs.
Access Code 13
Resolution 1
Minimum Value 1
Maximum Value 99

Table of Contents

Other manuals for Tuttnauer 3870 EHS

Related product manuals