15
The chamber pitch now needs to be adjusted
Measure to the proper amount of distilled water for the appropriate
model unit as listed below (the tolerances on these amounts are +2
oz. and –0 oz)
1730 2340/2540 3140 3850 3870
10-12
oz
300-350
ml
12-15
oz
350-450
ml
14-16
oz
420-480
ml
20-23
oz
600-690
ml
24-27
oz
720-810
ml.
Pour the proper amount of water into the chamber through the front
door of the unit (4)
This water should cover the bottom of the chamber to within +/- ½
inch of the groove in the front. (1)
If necessary adjust the front leveling feet, up or down, so that the
water lays in the chamber correctly. (2)
Once the chamber pitch adjustment is completed, the unit is ready
to operate
Note: It is imperative to have the correct amount of water in the
chamber each cycle for proper operation of the autoclave.
4.3 Lifting and carrying
Caution:
Before moving the autoclave, Make sure that the electric cord is
disconnected from the power and there is no pressure in the
chamber.
1. Disconnect the power supply cord.
2. Drain the water from the reservoir and vessel.
To avoid injuries, lifting and carrying should be done by two people.
Do not drop this device!