Corrections
10.1 Check section 7 for correct operating
procedures.
10.2 Make sure no items are touching the walls
of the chamber.
10.3 Set the sterilization time according to the
table in section 7.
10.4 Sterilize fewer towels or gowns or adjust
the front of the unit higher to allow more
water into the chamber. See installation
instructions section 4.2.
10.5 Tighten the door more, if leaking persists
then replace the door gasket.
10.6 Door bellows is leaking, have a technician
replace the bellows.
10.7 Follow procedure in section 8.5 if leaking
persists have technician replace the Safety
valve.
10.8 Have a technician check the machine.
11.1 Follow instruction in section 10.31
12.1 Apply 1 or 2 drops of 3 in 1 oil or some
white lithium grease to the screw and
bearing of the closing device see sec
8.1.2.4.
Possible cause check-up and tests
10.1 Chamber was not filled with the correct
amount of water.
10.2 Items are lying up against the chamber.
10.3 The sterilization time is set too high.
.
10.4 Items being sterilized are absorbing all
available steam (cloth towels or gowns).
10.5 Door gasket is leaking steam.
10.6 Steam is leaking at the closing device.
10.7 Safety Valve is leaking.
10.8 There is an internal steam leak or
electrical problem.
11.1 Door bellow locking pin is trapped.
12.1 Closing device needs oil.
Symptom
10. Items in the
chamber are
burning or melting.
11. Closing device does
not open when
pressure gauge
reads 0 psi.
12. Closing device is
hard to close.