39
Corrections
17.1 Open drain valve completely and blow
compressed air into the valve. This should
blow the clog back into the reservoir
where it can be removed. Cover the
reservoir opening to avoid splatter.
18.1 Open drain valve completely and blow
compressed air into the valve. This should
blow the clog back into the reservoir
where it can be removed. Cover the
reservoir opening to avoid splatter.
18.2 Have a technician replace the o-rings.
19.1 Follow cleaning instructions using
Chamber Brite section 9.
20.1 Use 100% steam distilled water.
20.2 Follow cleaning instructions using
Chamber Brite section 9.
21.1 Have a technician replace the door
bellows.
22.1 Tighten the door more.
22.2 Replace the door gasket (see sec. 8.4.
23.1 Have a technician replace the multi-
purpose valve.
Possible cause check-up and tests
17.1 Drain valve is clogged.
18.1 Debris is stopping the valve from seating
properly.
18.2 Drain valve o-rings are worn.
19.1 Unit has not been cleaned recently.
20.1 Not using distilled water.
20.2 Unit has not been clean recently.
21.1 The door bellows is leaking.
22.1 Door may not be tightened down enough.
22.2 Door gasket may be worn or cracked.
23.1 The internal spring in the multi-purpose
valve has broken.
Symptom
17. Reservoir does not
drain
18. Drain valve is
leaking
19. Chamber is black or
dirty looking
20. Chamber has a water
mark along bottom
21. Steam escaping from
the closing device
area.
22. Steam leaking from
the door seal.
23. Multi-purpose valve
turns backwards.