41
Corrections
31.1
1. a. Turn the Multi-purpose valve to
EXH. & DRY.
b. Turn the handle of the Door Closing
Device ¼ of a turn in the clockwise
(closing direction). The pin will be
released, enabling the operator to
open the door.
2. If this does not correct the problem then
a Technician will need to be called
32.1
a. A Tuttnauer Pouch Rack is
recommended to provide proper
spacing of bagged instruments. This
will allow for better exposed to steam
for sterilization and better circulation
of air for drying.
b. If a Pouch Rack is not used them make
sure the instruments on the tray are only
one level deep and that the paper /
plastic bags are plastic side down.
32.2 End cycle according to para. 7.
Possible cause check-up and tests
31.1 Locking pin is trapped.
32.1 Bagged instruments are placed too close
to each other.
32.2 Cycle ending has not been performed
correctly.
Symptom
31. Door handle cannot
be turned counter-
clockwise for the
door to be opened
and there be no
pressure in the.
32. Wrapped or bagged
instruments are not
drying properly (see
also sec 7.11).