3.2 Parameter
Character data and string data
Character data is consisted of ASCII characters. String data is consisted of ASCII charcters with quotation mark
(“”).
Numercial data
Numerical data includes integer number (NR1), Fixed point number (NR2), and Floating point number (NR3).
The numerical data range is ±9.9E37.
Examples of NR1:
Examples of NR2:
Examples of NR3:
123
12.3
12.3E+5
+123
+1.234
123.4E-56
-123
-123.4
3.3 Command System
3.3.1 SPEED
SPEED?
SPEED
Function:
Query or set test speed.
Parameter:
FAST | MEDium | SLOW
Return:
FAST | MEDium | SLOW
3.3.2 DISPlay
:DISPlay:PAGE?
:DISPlay:PAGE
Function:
Query or set display page.
Parameter:
MEASurement
BNUMber
MSETup
SYSTem
(Measurement display)
(Comparator function)
(Measurement setup)
(System setup)
Return:
< MEAS DISP > | < BIN DISP > | < MEAS SETUP > | < SYSTEM SETUP > | < ABOUT > |
< INTER FILE LIST >
:DISPlay:RFONt
Function:
Query or set font size. Query or set display ON/OFF.
Parameter:
LARGe
TINY
OFF
ON
(Large font size)
(Small font size)
(Measurement data display off)
(Measurement data display on)
Return:
OFF | LARGE | TINY
:DISPlay
Function:
Query or set measurement data display mode.
Parameter:
DIRect (Direct reading)
PERcent (Percentage deviation)
ABSolute (Absolute deviation)
Return:
DIRECT | PERCENT | ABSOLUTE