- 32 -
: COMParator:COUNt:CLEar
Function:
Clear counter number of ll ranges.
3.3.12 LIMit
: LIMit:NOMinal
Function:
Query or set nominal value.
Parameter:
<value>
Return:
<value>
<value> is NR3 format.
: LIMit:BIN<n> <HIGH>,<LOW>
Function:
Query or set high and low limits of each range.
Parameter:
n (1~3, in NR1 format.)
high, low (-100~100, in NR3 format.)
Return:
<HIGH>,<LOW>
Note:
If there is no setting for high and low limits, the return value will be “9.9999e+37”.
: LIMit:SECondary <HIGH>,<LOW>
Function:
Query or set high and low limits of secondary parameters.
Parameter:
high, low (in NR3 format.)
Return:
<HIGH>,<LOW>
Note:
If there is no setting for high and low limits, the return value will be “9.9999e+37”.
3.3.13 HANDLER
:HANDler:MODE
Function:
Query or set comparator output mode.
Parameter:
CLEAr | HOLD | PULSe
Return:
CLEAR | HOLD | PULSE
: HANDler:PULSe
Function:
Query or set pulse width.
Parameter:
<numeric_value> | MIN | MAX
Return:
<numeric_value>
Note:
<numeric_value> is 1~9999. Unit is default at ms.
: HANDler: EDGE
Function:
Query or set trigger edge.
Parameter:
RISing | FALLing
Return:
RISING | FALLING