EasyManuals Logo

twintex LCR Series User Manual

twintex LCR Series
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
- 31 -
3.3.9 TRIGger
: TRIGger
Function:
Query or set trigger mode. Or immediately trigger for one time.
Parameter:
INTernal (Internal trigger)
EXTernal (External trigger)
IMMediate (Immediately trigger for one time)
Return:
INTERNAL | EXTERNAL
: TRIGger:DELay
Function:
Query or set trigger delay time.
Parameter:
0~6000 (ms)
Return:
0~6000
3.3.10 CORRection
: CORRection
Function:
Clear data.
Parameter:
OPEN (Single point open circuit clearing)
OPEN_ALL (All points open circuit clearing)
SHORt (Single point short circuit clearing)
SHORt_ALL (All points short circuit clearing)
3.3.11 COMParator
: COMParator[:STATe]
Function:
Query or set comparator status.
Parameter:
ON | OFF | 1 | 0
Return:
1 | 0
: COMParator:Auxiliary BIN
Function:
Query or set auxiliary status.
Parameter:
ON | OFF | 1 | 0
Return:
1 | 0
: COMParator:BIN CLEar
Function:
Clear all upper and lower limits.
Parameter:
ON | OFF | 1 | 0
Return:
1 | 0
: COMParator:COUNt[:STATe]
Function:
Query or set counter status.
Parameter:
ON | OFF | 1 | 0
Return:
1 | 0
: COMParator:COUNt:DATA?
Function:
Query counter number of each range.
Return:
<count numbers of fail range>,<counter number of range 1>,<counter number of range
2>,<counter number of range 3>,<counter number of auxiliary range>

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the twintex LCR Series and is the answer not in the manual?

twintex LCR Series Specifications

General IconGeneral
Brandtwintex
ModelLCR Series
CategoryMeasuring Instruments
LanguageEnglish