EN 25
0,1
0,37
0,11
0,22
1
1,2
0,35
0,7
10
3,7
1,1
2,2
100
12
3,5
7
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTES 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTES 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Table 7: Recommended separation distances between portable and mobile RF
communications equipment and OMEGO® Basic system
2.1.2 Usage area
Device usage is principally limited to clean, dry interiors in professional health care
establishments.
Operation
Temperature: 10 ... 30 °C
Humidity: 30 ... 75 % relative humidity
Air pressure: 70 kPa 106 kPa
Storage und transport
Temperature: -20 ... 60 °C
Humidity: 20 ... 90 % relative humidity. No dew
Air pressure: 70 kPa … 106 kPa
Caution:
The OMEGO
®
Basic system may not be used in explosion-prone zones AP and APG
according to EN 60601-1/2006+A1/2013.