2.6 Monthly functionality check/Recurring check
2.6.1 Functionality checklist
The functionality check described here must be performed every month. Perform the
check even if OMEGO
®
Basic indicates a malfunction (e.g. in case of unusual sounds,
jolting movements, elementary damages etc.). The person responsible for the check
must be a certified OMEGO
®
Basic user.
To be checked:
Malfunction:
To do:
Protective casing
Casings are unstable
Casings are missing
Mounting bolts are missing or
loose
Casings are damaged
Do not perform any further
therapies
Contact TYROMOTION GmbH
Visible external
deformations
Parts are deformed
Parts are asymentrical
Parts are faulty
Do not perform any further
therapies
Contact TYROMOTION GmbH
Emergency
off-switch
Activating emergency off-
switch does not trigger
emergency stop
Do not perform any further
therapies
Contact TYROMOTION GmbH
Cleaning
Handles are dirty
OMEGO
®
Basic is dirty
Foot rests are very dirty
Do not perform any further
therapies
Clean the dirty parts as described
in chaper 4.5.2
Locking
Monitor handle bracket locking
mechanism is faulty
Foot rest locking mechanism is
faulty
Chair safety catch locking
mechanism is faulty
Do not perform any further
therapies
Contact TYROMOTION GmbH
kontaktieren
Stability
Device skids increasingly during
therapy
Do not perform any further
therapies
Check the anti-skid pads on the
device’s underside
Contact TYROMOTION GmbH
Table 9: Checklist