2.6 Monthly functionality check/Recurring check
2.6.1 Functionality checklist
The functionality check described here must be performed every month. Perform the
check even if OMEGO
®
Basic indicates a malfunction (e.g. in case of unusual sounds,
jolting movements, elementary damages etc.). The person responsible for the check
must be a certified OMEGO
®
Basic user.
To be checked:
Malfunction:
To do:
Protective casing
• Casings are unstable
• Casings are missing
• Mounting bolts are missing or
loose
• Casings are damaged
• Do not perform any further
therapies
• Contact TYROMOTION GmbH
Visible external
deformations
• Parts are deformed
• Parts are asymentrical
• Parts are faulty
• Do not perform any further
therapies
• Contact TYROMOTION GmbH
Emergency
off-switch
• Activating emergency off-
switch does not trigger
emergency stop
• Do not perform any further
therapies
• Contact TYROMOTION GmbH
Cleaning
• Handles are dirty
• OMEGO
®
Basic is dirty
• Foot rests are very dirty
• Do not perform any further
therapies
• Clean the dirty parts as described
in chaper 4.5.2
Locking
• Monitor handle bracket locking
mechanism is faulty
• Foot rest locking mechanism is
faulty
• Chair safety catch locking
mechanism is faulty
• Do not perform any further
therapies
• Contact TYROMOTION GmbH
kontaktieren
Stability
• Device skids increasingly during
therapy
• Do not perform any further
therapies
• Check the anti-skid pads on the
device’s underside
• Contact TYROMOTION GmbH
Table 9: Checklist