EN 57
4.5.3 cleaning process
To minimize the risk of germ transmission, all surfaces and objects that are touched by
both the patient and the therapist should be periodically cleaned and disinfected. These
include e.g. also monitor, foot shells, bracket monitor handle, housing, etc.
1. If the patient perspires heavily during use, wipe the OMEGHandgrips dry
after the therapy before disinfecting them.
2. Moisten the disposable cloth, according to the product information leaflet, only
slightly with disinfectant. Wipe OMEGO® Hand Grips, Greaves, Foot Cups with
the clean, soft and lint-free cloth. Observe the contact time of the disinfectant
used according to the label and product information.
3. Depending on the disinfectant, it may be necessary to wipe the disinfected area
with water after the exposure time.
4. Dry the area wiped with water with a clean, non-abrasive, soft, lint-free cloth.
To clean the loops, proceed as follows:
1. Use a disinfectant spray (such as Bacillol® 30 Foam) to disinfect the fabric after
each treatment.
2. Clean the loops of the foot cup and the greaves weekly in the washing machine
at 40 ° C in the gentle cycle with the addition of a Hygienespülers be cleaned.
The use of a wash bag is recommended.