Usage of the device must be stopped immediately if one of the
malfunctions specified in the Table 9: Points for inspection occurs or
is suspected.
Casing damages (e.g. caused by transportation):
Never activate the device if the casing is damaged as this may have
unpredictable consequences (e.g. electric shock). Check each point
listed in the check list before activation of the device and contact
Tyromotion if necessary.
2.6.2 Recurring check
Recurring checks differ from the checks in chapter 2.6.1; the legislator may demand the
check described here while the checks in chapter 2.6.1 are intended, among other things,
to detect acute damage or wear of parts that necessitates replacement. The device
operator is responsible for performing both checks.
TYROMOTION GmbH has determined an interval of one year for recurring checks.
Recurring checks may only be carried out by professional and qualified personnel. The
device operator must ensure that the intervals for the recurring checks stipulated by him
are observed. Usage of the OMEGO
®
Basic system must cease if the inspection intervals
are not observed.
The recurring check must be performed according to EN 62353:2014.