EN – 50
4.3 Prior to training
4.3.1 Activating the OMEGO
®
Basic systems
The system can be activated after it has been properly connected. Activate the device
with the power switch. All components are now being supplied with electricity. Then turn
on the PC power switch to boot the PC. The PC status light indicates that the PC has been
turned on. The operating software can now be started after logging into Windows.
4.4 Application
The device settings have to be adjusted to the respective patient prior to his/her first
therapy session with the OMEGO
®
Basic.
Furthermore we recommend suggesting the patient washes and disinfects his/her
hands.
Choose the OMEGO
®
Basic in the tyroS menu “Device selsction”. The OMEGO
®
Basic will
then run a reference run. Once this is completed, the foot rests lock into the easy entry
position and the feet can be secured in the foot rests.
Only attachment systems provided by Tyromotion may be used to
attach the patient. Other attachment systems may pose a hazard.
Risk of injury by stuck clothes: The patient has always wear suitable,
tight clothes. Wide pants, long scarves and shawls, long hair etc. may
get caught in the pedals. The patient has not wear necklaces, jewelry
or similar. The patient has always wear closed footwear without laces.
All settings, especially maximum speed, maximum force and
maximum range of motion have to be determined with the treating
physician prior to the first therapy session.
During therapy, only trained personnel may be around the patient
and therapy system. Ideally, the therapist is right next to the patient
to instruct him/her during therapy.