Vaisala CARBOCAP GMW90 Series - Set Analog Output Scaling

Vaisala CARBOCAP GMW90 Series
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User's Guide _______________________________________________________________________
44 ___________________________________________________________________ M211659EN-C
Example (show current configuration):
>pass 9000
>amode
Aout 1 range (mA) : 4.00 ... 20.00 (error: 3.60)
Aout 2 range (mA) : 4.00 ... 20.00 (error: 3.60)
Aout 3 range (mA) : 4.00 ... 20.00 (error: 3.60)
Example (set channel 1 to 0 ... 20 mA output, with error level at 21 mA):
>amode 1 0 20 21
Aout 1 range (mA) : 0.00 ... 20.00 (error: 21.00)
>
Set Analog Output Scaling
Use the ASEL command to select the output parameter and scaling for
analog output channels.
ASEL [channel parameter lo_value hi_value]<cr>
where
channel
=
Analog output channel, 1 ... 3.
parameter
=
Parameter that is output on the channel.
Available parameters are:
CO2 carbon dioxide
RH relative humidity
T temperature
Tdf dew/frostpoint temperature
Td dewpoint temperature
Tw wetbulb temperature
h enthalpy
x mixing ratio
a absolute humidity
dTd dew/frostpoint depression
lo_value
=
Low limit of the scaling, in the units of the selected
parameter.
hi_value
=
High limit of the scaling in the units of the selected
parameter.
Example (set channel 1 to output dewpoint, in the range -10 ... 20 °C):
>pass 9000
>asel 1 TD -10 20
Aout 1 quantity : Td (-10.00 ... 20.00 'C)