Chapter 5 _______________________________________________________________ Maintenance
VAISALA ________________________________________________________________________ 67
Adjustment Using Display and Trimmers
CAUTION
point. If you wish to apply a
. Corrections applied using the
NOTE
User calibration settings (adjustment by trimmers or ser
vice port) are
stored in the
measurement modules. If you replace a module, there is no
need to undo previous adjustments
.
1. To enter the adjustment screen, open the slide and rotate the CO
2
,
T, or RH trimmer slightly during normal measurement. If the
trimmer is not centered, you see the trimmer centering screen first.
Simply turn the trimmer to the center and wait for the progress bar
to complete.
1111-075
Figure 26 Trimmer Centering Screen
2. In the adjustment screen, turn the trimmer to set the desired
correction. To commit the change, stop turning the trimmer and
wait.
1310-092
Figure 27 Adjustment Screen
3. If you have adjusted the CO
2
measurement, wait for three minutes
for the reading to stabilize.
4. Verify the adjusted measurement from measurement screen.