Chapter 5 _______________________________________________________________ Maintenance
VAISALA ________________________________________________________________________ 65
Effect of Pressure on CO
2
Measurement
Ambient pressure affects the CO
2
reading of GMW90. If the ambient
pressure differs from the normal pressure 1013 hPa, the transmitter can
correct the reading if you enter the correct pressure value to the
transmitter using an MI70 indicator, computer (ENV command), or using
Modbus protocol.
If you want to apply the pressure compensation in the building control
system instead, multiply the measured CO
2
value using the appropriate
compensation multiplier from Table 11.
Table 11 Effect of Elevation on CO
2
Measurement
Elevation (m)
Elevation (ft)
Pressure (hPa)
Compensation
Multiplier
0
0
1013
1
100
328
1001
1.017
200
656
989
1.034
300
984
978
1.051
400
1312
966
1.067
500
1640
955
1.084
600
1969
943
1.100
700
2297
932
1.116
800
2625
921
1.132
900
2953
910
1.148
1000
3281
899
1.164
1100
3609
888
1.179
1200
3937
877
1.195
1300
4265
867
1.210
1400
4593
856
1.225
1500
4921
846
1.240
1600
5249
835
1.255
1700
5577
825
1.269
1800
5906
815
1.284
1900
6234
805
1.298
2000
6562
795
1.312
2100
6890
785
1.326
2200
7218
775
1.340
2300
7546
766
1.354
2400
7874
756
1.368

Table of Contents