Alarm Function
The MI70 indicator can produce an alarm when a measured quantity is 
outside the allowed range. The range is controlled by setting the lower 
and upper limits. When the alarm is active, the indicator will show an 
alarm message, play a beeping sound, and blink the display backlight. 
The alarm can only be set for one quantity at a time. Set the alarm as 
follows:
1. Open the Functions menu.
2. Select Alarm.
3. Select the measured Quantity for the alarm. Only quantities that 
are currently measured can be selected.
4. Set two limit values for the alarm using the Limit options.
5. Check the Alarm on/off checkbox to enable the alarm. After you 
enable the alarm and exit the menu, the alarm symbol will appear 
on the display, and flash briefly.
0812-019
Figure 17 Alarm Menu
NOTE
The alarm will automatically turn off if you change the measurement 
configuration (change the measured quantities, or connect/disconnect 
devices). If you use the alarm function, always check the alarm 
settings after changes.