EasyManua.ls Logo

Vaisala PTB330TS - 1-Point Adjustment (LCP1); Figure 21 Inside of Open Barometer

Vaisala PTB330TS
89 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 ______________________________________________________________ Maintenance
VAISALA_______________________________________________________________________ 47
0705-091
Figure 21 Inside of Open Barometer
1-Point Adjustment (LCP1)
The LCP1 command performs linear correction for barometer
module/module P1. Use the command LCP1
- to activate or deactivate the linear adjustment function
- to enter new linear offset and offset/gain pressure corrections to the
transmitter
- to edit existing linear offset and offset/gain pressure corrections.
Numbers refer to Figure 21 on page 47:
1 = Adjustment button with indicator LED
2 = Power supply module (not included in PTB330TS)
3 = Power supply mode selections (do not change)
4 = Service port (RS-232)
5 = Module 1/Module 3 connectors
6 = User port (wired to the M12 connector)
7 = Module 2/Module 4 connectors

Table of Contents