Chapter 6 ______________________________ Serial Line Communication With PTB330 Barometer
VAISALA_______________________________________________________________________ 69
Example:
>r
986.62
986.63
986.63
986.63
...
S
Use the S command to end the RUN mode. After this command all other
commands can be used. You can also press the Esc key on your
keyboard or reset the transmitter to stop the outputting.
SEND
Use the SEND command to output the reading once in STOP mode. The
output format can be adjusted with the FORM command.
Example:
>send
986.61
>
Measurement Settings
AVRG
Use the AVRG command to set the barometer measurement averaging
time. Valid range of averaging time is 1 ... 600 seconds.
AVRG [x]
where
x = Averaging time in seconds