Chapter 6 ______________________________ Serial Line Communication With PTB330 Barometer
VAISALA_______________________________________________________________________ 71
HHCP
The HHCP command shows or sets the altitude for height corrected
pressure. The valid range of HCP height is -30...+30 m.
Example:
>hhcp
HCP height : 0.0 m ?
>
HQFE
The HQFE command displays or sets altitude for QFE corrected
pressure. The valid range of QFE height is -30...+30 m.
Example:
>hqfe
QFE height : 0.0 m ?
>
HQNH
The HQNH command displays or sets altitude for QNH corrected
pressure. The valid range of QNH height is -30...+3000 m.
Example:
>hqnh
QNH height : 0.0 m ?
>
Formatting Measurement Output
FORM
Use the serial line command FORM to change the output format and
quantities for serial line use. See Table 3 on page 16 for available
quantities.