User's Guide ______________________________________________________________________
66 ___________________________________________________________________M210983EN-A
A female M12 connector with screw terminals is provided with the
PTB330TS; see Figure 27 on page 66.
0812-018
Figure 27 Female M12 Connector with Screw Terminals
List of PTB330 Serial Commands
The bold text in the brackets is the default setting. To issue a command,
type it on your terminal application and press the Enter key.
Measurement commands
INTV [0...255 s/min/h/d] Shows or sets the continuous output interval (for RUN mode).
R Changes the serial mode to RUN and starts outputting
measurement results according to the FORM string (with
interval defined by INTV).
S Changes the serial mode to STOP and stops outputting
measurement results.
SEND Shows the measurement results according to the configured
form.
Measurement setting commands
AV RG [1...600 s] Set the barometer measurement averaging time (in seconds)
DPMAX Shows or sets the maximum allowed pressure difference
between barometer modules.
TQFE [-80...+200 °C] Shows or sets the temperature for QFE corrected pressure.
HHCP [-30...+30 m] Shows or sets the altitude for height corrected pressure.
HQFE [-30...+30 m] Shows or sets the altitude for the QFE corrected pressure.
HQNH [-30...+3000 m] Shows or sets the altitude for the QNH corrected pressure.