1.4 How the Vapro Works
A 10 microliter specimen is aspirated into a micropipettor tip. The
specimen is then inoculated into a solute-free paper disc in the
sample holder, whereupon the sample holder is pushed into the
instrument and the sample chamber is locked. Locking initiates the
automatic measurement sequence.
The sensing element is a fine-wire thermocouple hygrometer. This
is suspended in a unique, all-metal mount, which when joined with
the sample holder, forms a small chamber enclosing the specimen.
As vapor pressure equilibrates in the chamber airspace, the thermo-
couple senses the ambient air temperature, thus establishing the
reference point for the measurement. Under electronic control, the
thermocouple then seeks the dew point temperature within the
enclosed space, giving an output proportional to the differential tem-
perature.
The difference between the ambient temperature and the dew point
temperature is the dew point temperature depression–an explicit
function of solution vapor pressure.
Dew point temperature depression is measured with a resolution of
0.00031 °C. The microprocessor-controlled measurement cycle
requires 80 seconds.
Appendix C contains the theory of operation of the vapor pressure
osmometer.