APPENDIX
G
Setup Menu
95
The Setup Menu allows you to select available lan-
guages: English, French, or German and available
measurement units: mmol/kg or kilopascals at 25 °C
(= -2.5 x mmol/kg). It also allows you to run a Self
Test on the instrument to check basic input and out-
put functions.
Language and Units of Measurement are initially
designated at the factory and stored in non-volatile
memory. They become default settings on power up
of the osmometer and are displayed briefly at that
time.
To change these settings:
1
Turn off power. Wait approximately 10 seconds.
2
While pressing both the SELECT and ENTER
keys, turn on the power. Wait several seconds
while the display shows the Wescor logo,
default selections, and finally the Setup Menu.
3
Select desired language or unit of measure-
ment using the SELECT button to move the point-
er to your choice. Press ENTER to add selection
to memory.
4
Once you have selected your preferences you can
exit the Setup Menu. The instrument defaults to
these settings until they are changed.