SECTION 4
PREVENTIVE MAINTENANCE
4.4 Reinstalling the TC Head
5
Replace the access cover.
6
Close the sample chamber.
7
Turn on the power. Allow the instrument to complete the initial-
ization sequence and reach thermal equilibrium (see Section
4.5) before attempting to calibrate the osmometer.
8
Calibrate using the 290, 1000, and then 100 mmol/kg stan-
dards. After calibrating using the 100 mmol/kg standard, select
CLEAN TEST and then press ENTER.
9
If the Clean Test reveals contamination, refer to Section 4.6 and
(if necessary) Section 5.1.
52