4.7.2 Emergency Stop Switch
During operation, the following emergencies may occur:
X-ray emission even after the Exposure Switch has been released
Physical injury to the patient or damage to the equipment
Other emergencies
If a problem occurs during image acquisition, press the red Emergency Stop Switch
to immediately stop the moving parts and cut off all power to the equipment’s electrical
components. To restart the equipment, turn the Emergency Stop Switch clockwise
until it pops up.
The Emergency Stop Switch is located under the bottom of the Handle Frame.
4.7.3 Exposure Switch
The Exposure Switch allows the operator to control image acquisition from outside of
the X-ray room.
Press and hold the Exposure Switch down until acquisition is completed. Premature
release of the Exposure Switch will abort image acquisition.
Pressing the Exposure Switch activates the LED indicator to turn yellow. This color
indicates that the X-ray is being emitted.
The Exposure Switch is detachable. Ensure the Exposure
Switch cable is not detached from the unit accidentally
during the operation.
Keep vocal/visual contact with the patient during exposure. If
any problem occurs during exposure, release the Exposure
Switch immediately.

Table of Contents

Other manuals for Vatech A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech A9 and is the answer not in the manual?

Vatech A9 Specifications

General IconGeneral
BrandVatech
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals