EasyManua.ls Logo

Vatech PaX-i Insight

Vatech PaX-i Insight
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
136 PaX-i Plus / Insight (Model: PCH-30CS) Installation Manual
4. Before Installation:
Site Check List
Yes
No
Is the room large enough?
At minimum,
with CEPH unit: 2,805 mm (L) x mm 2,233 (W) x 2,420 mm (H)
/ 111x 88x 96”
Without CEPH unit: 1,890 mm (L) x 2,233 mm (W) x 2,420 mm
(H) / 75” x 88x 96
Is the door entrance wider than 800mm (32”)?
Is a radiation protection plan in place?
Do equipment and PC use the same dedicated circuit?
Does the electrical input condition to the installation site meet
the MEIGaN requirements?
Is the local Network IP address of clinic 192.168.33.xx?
Is a compressor or air conditioner suction located right next to
X-ray Room?
Is the floor flat and level?
Is the carpet on the floor? If so, remove it.
Before Opening Boxes
Yes
No
Did the delivery company carry and handle with caution?
Did the installers take pictures of boxes before opening?
Did the installer make sure there are not any suspicious holes
or scratches on the box?
Is the ShockWatch indicator red?
Is the TiltWatch indicator red?

Table of Contents

Related product manuals