9 Diagnostics and servicing
MINITRAC 31 • Foundation Fieldbus
41782-EN-211203
This status message is inactive by default.
Code
Text message
Cause Rectication
F008
Error multi sensor commu-
nication
Additionalsensorsnotswitchedon
EMCinuences
No other sensor available
Checkwiringbetweenthesensors
Connect the sensors correctly and make
them ready for operation
F013
Sensor signals a fault
Erroronthecurrentinput/digitalinput
No valid measured value
Connectedinstrumentswithoutfunction
Check current input
Check connected instruments (Second-
aryinstrument)
F016
Adjustment data exchanged
Valuesofthemin.andmax.adjustment
exchanged
Correct adjustment data
F017
Adjustment span too small
The values of the min. and max. adjust-
ment are too close together
Correct adjustment data
F025
Invalidlinearizationtable
Wrongoremptylinearizationtable
(1074,1075,1080,1100,1106)
Wrongvalueinthelinearizationtable
(1143,1144)
Createlinearizationtable
Correctlinearizationtable
F030
Process value out of limits
Processvaluesarenotwithintheadjust-
ed measuring range
Repeat adjustment
F034
EPROMhardwareerror
Electronics defective Restart instrument
Exchanging the electronics
F035
EPROM data error
Error in the internal instrument commu-
nication
Carry out a reset
Exchanging the electronics
F036
Faulty program memory
Errorduringsoftwareupdate Repeatsoftwareupdate
Exchanging the electronics
F037
RAMhardwareerror
Error in RAM Restart instrument
Exchanging the electronics
F038
Secondary signals failure
Connection cable to the Secondary in-
strument interrupted
InstrumentnotdenedasSecondaryin-
strument
One of the Secondary instruments sig-
nals a failure
Check the connection cable to the Sec-
ondary instrument
DeneinstrumentasSecondary
Check Secondary instruments
F040
Hardwareerror
Instrumentdefective(1092,1126)
Temperatureoutsidethespecication
(1091)
Restart instrument
Exchanging the electronics
Cooltheinstrumentorprotectitwithiso-
lationmaterialagainstheat/cold
F041
Photomultiplier error
Error in the measured value recording Restart instrument
Exchanging the electronics
F052
Faultyconguration
Invalid parameter adjustment Carry out a reset
Failure