9 Diagnostics and servicing
MINITRAC 31 • Foundation Fieldbus
41782-EN-211203
Code
Text message
Cause Rectication
F066
Faulty adjustment
Adjustment not yet carried out
Errorduringadjustmentorwhenenter-
ing the linearisation table
Carry out adjustment
Carry out linearisation
F068
Count rate too high
Faultyinstrumentsettings(1031)
Faulty steam density compensation
(1101)
Carry out a reset
Check Secondary instrument (steam
density)
F072
Limit exceeded
Faulty instrument settings Carry out a reset
F073
Error real value correction
Faulty real value correction Carry out real value correction again
F080
System error
Instrument error Restart instrument
Call our service
F086
Communication error
Error in the Fieldbus communication Restart instrument
Call our service
F114
Error real time clock
Discharge accumulator Readjust real time clock
F120
Filter time error
Faulty or missing instrument adjustment Carry out adjustment
F121
Faulty participant list on the
multisensor communica-
tion bus
Secondary instruments not found
Secondaryinstrumentwithwrongad-
dress
Check Secondary instruments
Check Secondary list in Primary instru-
ment
Correct address of Secondary instru-
ments
F122
Double addresses on the
multisensor communica-
tion bus
Instrumentaddressewasassignedsev-
eral times
Change instrument addresses
F123
X-ray alarm
External instruments cause radiation
Radiation above the max. adjustment
value
Determine reason for X-ray alarm
In case of brief X-ray radiation: Monitor
switchingoutputsforthistimemanually
F124
Alarm due to increased ra-
diation
Radiation dose too high Determine reason for increased radia-
tion
F125
Ambient temperature too
high
Ambient temperature on the housing
outsidethespecication
Cooltheinstrumentorprotectitwithiso-
lating material against radiation heat
F126
Error in the trend recording
Instrument error Call our service
F141
Communication error on the
multisensor communica-
tion bus
Secondaryinstrumentdoesnotanswer Check Secondary instruments
Tab. 4: Error codes and text messages, information on causes as well as corrective measures

Table of Contents

Other manuals for Vega MINITRAC 31

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vega MINITRAC 31 and is the answer not in the manual?

Vega MINITRAC 31 Specifications

General IconGeneral
BrandVega
ModelMINITRAC 31
CategoryAccessories
LanguageEnglish

Related product manuals