User Manual MicroDispensing Systems MDS 3010A-FD and MDS 3020A-FD
User Manual MicroDispensing Systems MDS 3010A-FD and MDS 3020A-FD
VTK-TS-MA-032 Revision D 61
Select the menu item “Calibration H(eater)1” by pressing the enter
button.
Select the service code by using the arrow keys and confirm with the
enter button.
Use the arrow keys until the menu item “Value Measurement” is shown
on the display.
To confirm “Value Measurement” press enter.
To confirm “Measure Values? Press enter” press the enter button.
Make sure that the thermocouple (type K) is connected. If not, wait 5 min
after connecting. After that confirm the prompt “Put in a Type K
Thermocouple!” by using the enter button.
Put the thermocouple into ice water. The ice water should contain many
little pieces of broken ice. Wait for 30 s.
Then confirm the message “Put the Thermo(couple) into Icewater“.
Place the thermocouple into boiling water, but it should not touch the
ground. The water should be boiling only a little to avoid steam bubbles
caused by the higher temperature. Wait for 30 s and after that confirm
the message “Put the Thermo(couple) into boiling W(ater)” by pressing
the enter button.