INSTALLATION
2-5
VersaPulse Select Service Manual
0621-499-01 12/95
®
®
11.) Add coolant.
With the system turned on, top off the fill reservoir. Cycle the machine off and on several times,
adding coolant as necessary until the fill reservoir level stays up to approximately half full as the
system runs.
12.) Operational and Safety Checkout.
Attach a test fiber, remote interlock plug, and footswitch, then check system operation as follows:
a.) Footswitch, remote control (if included) and remote interlock plugs connect properly to jacks.
b.) Fiber attaches to the fiber port.
c.) Touch screen display matrix is operating properly.
d.) Touch screen inputs are functional.
e.) Remote control (if installed) display and input buttons are functional.
f.) Aiming beam responds to off, low, medium, and high settings.
CAUTION: The remaining operational tests include firing the system and attempts to fire the system
with one disabling condition. Select a low power setting and direct the test fiber output into a
calibrated power meter.
g.) Fire into the meter for several seconds, noting the power reading on the meter.
i.) Confirm that the total energy display value has incremented, then confirm that the value clears
back to zero when RESET is touched.
j.) Disconnect the footswitch and confirm that the "ATTACH FOOTSWITCH" message appears on the
display.
k.) Reconnect the footswitch, then disconnect the fiber and confirm that an "ATTACH FIBER"
message appears on the display. Depress the footswitch and confirm the system will not fire.
l.) Reconnect the fiber, then disconnect the remote interlock plug and confirm the " REMOTE
INTERLOCK" message appears on the display.
m.) Reconnect the remote interlock plug, then go to STANDBY. Depress the footswitch and confirm
the system will not fire.
n.) Pull the blast shield out, then confirm the "REPLACE BLAST SHIELD" message appears on the
display. Depress the footswitch and confirm the system will not fire.
o.) Reinsert the blast shield.